Real-time gas analysis. Wherever the process is.
* QMS · Compact Field-Type Gas Analysis System
iquad-s is a compact, portable QMS gas analysis system designed for on-site real-time measurement where sampling-then-analysis is impractical.
It injects target gas directly for immediate analysis, and uses monoatomic tracer gas (Kr) concentration to calculate total volumetric flow rate — a method capable of measurements that FTIR cannot perform.
* Key Features
01
Direct Real-Time Injection
Gas is injected directly into the analyzer without a prior sampling step. Ideal for on-site measurement where sample collection is difficult — such as process exhausts and tight-space environments.
02
Kr-Based Flow Rate Calculation
Total volumetric flow rate is calculated by measuring the concentration of monoatomic Kr tracer gas — a capability unique to QMS that is not achievable with FTIR systems.
03
Compact & Portable
At 471×195×260 mm and 15 kg, iquad-s is engineered for field use. Module-type inlet system and LAN/RS-232-C communication enable flexible integration into existing measurement setups.
* Product Overview
Field-Deployable QMS for Real-Time Gas Analysis
iquad-s is a field-deployable QMS system designed for direct, real-time gas measurement at the process point. Unlike laboratory-based analysis, it operates without a separate
sampling step, making it ideal for environments where sample collection is difficult or impractical. Volumetric flow rate is calculated using Kr tracer gas concentration, enabling
accurate process monitoring. The system is widely deployed for GHG reduction facility efficiency measurement and has been supplied to Samsung Electronics, GST, Ecopro,
and other leading industrial facilities.
System Overview
Model / Principle
iquad-s / QMS (Quadrupole Mass Spectrometry)
Detector
Faraday Cup & Secondary Electron Multiplier
Filament
Yttriated Iridium or Tungsten
Inlet Type
Module type inlet system
Performance
Mass Range
0–200 amu (Max. 1–500 amu)
Detection Limit
5 ppm ±3% (at N₂)
Resolution
1 amu
Min. Detection Partial Pressure
2×10⁻¹² mbar
Sensitivity
6×10⁻⁴ A/mbar (for Ar)
Measurable Gases
H₂, CH₄, CO, CO₂, N₂, O₂, He, Ne, Ar, Xe, Kr, SF₆, CF₄, NF₃, etc.
System Configuration
Main Power
AC 220V, 12.5A, 50/60 Hz, Single-phase
Operating Temp.
RT ~ 40°C
Humidity
0–80% (non-condensing)
Communication
RS-232-C / LAN
Output
iquad-s software
Installation Requirements
Dimensions (W×D×H)
471 × 195 × 260 mm
Weight
15 kg
* Applications
01
GHG Scrubber Efficiency
Standard measurement system for semiconductor/display GHG reduction device (scrubber) efficiency under Korea's Emissions Trading Scheme (ES 13501, KSI 0587:2023).
02
On-Site Process Exhaust
Direct measurement of process exhaust gas composition at the source — without transport or sample container, preserving sample integrity.
03
Flow Rate Measurement
Volumetric flow calculation using Kr tracer gas concentration for process streams where conventional flow meters cannot be used.
04
Field Sampling Campaigns
Mobile gas analysis for industrial sites, environmental surveys, and confined-space assessments where a fixed laboratory instrument is not available.