Every gas species,
measured with precision.
* QMS · General-Purpose Gas Analysis System
iquad is a Quadrupole Mass Spectrometry (QMS) gas analysis system designed for reliable multi-component measurement in laboratory environments.
With fully automated software, it delivers accurate quantitative data without requiring expert operators — from out-gassing tests to sealed OLED cell analysis.
* Key Features
01
Fully Automated SW
el's proprietary automation software handles calibration, measurement, and result output without manual steps — reducing operator error and delivering consistent data quality across all users.
02
Multi-Component Simultaneous
A single measurement cycle identifies and quantifies all gas species present in the sample — no sequential scanning required. Detection limit: 5 ppm ±3% (at N₂), 1 amu resolution.
03
Wide Mass Range
Standard mass range 0–200 amu (expandable to 1–500 amu). Covers light gases (H₂, He) through heavy species (SF₆, Xe), with both Faraday Cup and Secondary Electron Multiplier detectors.
* Product Overview
Laboratory-Grade QMS System
iquad (formerly isepa) is el's stationary, laboratory-grade QMS gas analysis system. It simultaneously identifies and quantifies multiple gas species
— including O₂, N₂, H₂, Ar, SF₆, CF₄, NF₃, and others — using a Quadrupole Mass Spectrometer operating in the low-vacuum range. The fully automated software minimizes human intervention,
reducing measurement error and enabling consistent, high-quality analytical results. Batch-type inlet design allows collected samples to be analyzed with high repeatability.
System Overview
Model / Principle
iquad / QMS (Quadrupole Mass Spectrometry)
Detector
Faraday Cup & Secondary Electron Multiplier
Filament
Yttriated Iridium or Tungsten
Inlet Type
Batch type (volume diffusion method)
Mass Filter
1 ea
Performance
Mass Range
0–200 amu (Max. 1–500 amu)
Detection Limit
5 ppm ±3% (at N₂)
Resolution
1 amu
Measurable Gases
H₂, CH₄, CO, CO₂, N₂, O₂, He, Ne, Ar, Xe, Kr, SF₆, CF₄, NF₃, etc.
System Configuration
OS / Control
Windows 10 / Touch screen · Auto & manual control program
Main Power
AC 220V, 12.5A, 50/60 Hz, Single-phase
Operating Temp.
RT ~ 40°C
Humidity
0–80% (non-condensing)
Installation Requirements
Dimensions (W×D×H)
950 × 780 × 1,500 mm
Weight
200 kg
Min. Operating Space
1,100 × 1,400 × 1,800 mm
Utilities
Air compressor (5–8 kgf/cm²) & N₂ standard gas
* Applications
01
Out-Gassing Analysis
Characterization of gases released from materials, components, or packaged devices under vacuum or elevated temperature conditions.
02
OLED Cell Internal Gas
Absolute quantification of internal atmosphere inside sealed OLED panels — identifying contaminants, moisture, and organic outgassing species.
03
Vacuum System QC
Verification of vacuum chamber cleanliness and residual gas composition before and after process runs.
04
Lab Gas Analysis Service
General-purpose qualitative and quantitative gas analysis for customer-submitted samples across semiconductor, display, steel, and research sectors.